Specifications:

Style: FD-1
The tester is designed and manufacturer according to YY0267-2016 5.5.10 <<Hemodialysis and Hemopurification Device External Circle Blood Line>> It applies external blood line examination

1)、Flow range at 50ml/min ~ 600ml/min
2)、Accuracy: 0.2%
3)、Negative pressure range: -33.3kPa-0kPa;
4)、High precise mass flowmeter installed;
5)、Thermostatic water bath installed;
6)、Keep constant negative pressure
7)、Testing result automatically printed
8)、Real-time display for error range

Product Structure

This device is composed of water bath box, high precision linear step control pressure regulator, pressure sensor, high precision flow meter, PLC control module, automatic following servo peristaltic pump, immersion temperature sensor, switching power supply and so on.

A temperature and humidity sensor is installed outside the device to measure the ambient temperature and humidity.